![Keysight Technologies N8480 Series Operating And Service Manual Download Page 78](http://html.mh-extra.com/html/keysight-technologies/n8480-series/n8480-series_operating-and-service-manual_1962056078.webp)
3
Service
60
N8480 Series Power Sensors Operating and Service Guide
Diassembly/Reassembly Procedures
Diassembly Procedure
Disassemble the power sensor by performing the following steps:
•
Diassemble the power sensor only in a static free workstation.
Electrostatic discharge renders the power sensor inoperative.
•
At the rear of the power sensor, insert the blade of a screwdriver
between the plastic shells (See
). To prevent damage to the
plastic shells use a screwdriver blade as wide as the slot between the
two shells.
•
Pry alternately at both sides of the connector until the plastic shells are
apart. Remove the shells and the magnetic shields.
Figure 3-2
Removing power sensor shell
Reassembly Procedure
Replace the magnetic shields and the plastic shells. Snap the plastic shells
together.