Specifications and Characteristics
2
N8480 Series Power Sensors Operating and Service Guide
37
Settling Time
In normal and x2 mode, manual filter, 10 dB decreasing power step (not
across the switching point), the settling time is shown in
.
Figure 2-15
Autofilter, default resolution, 10 dB decreasing power step (not across the
switching point)
Table 2-7
Settling time in normal and x2 mode
Number of
Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time
(s)
(Normal mode)
0.15
0.2
0.3
0.5
1.1
1.9
3.4
6.6
13
27
57
Response Time
(s) (x2 mode)
0.15
0.18
0.22
0.35
0.55
1.1
1.9
3.5
6.9
14.5
33
x2 Mode
Maximum sensor power
Sensor
Dynamic
Range
180 ms
Minimum sensor power
400 ms
3.6 s
150 ms
Typical
Settling
Times
10 dB
10 dB
10 dB
20 dB
6.6 s
5 dB
Normal Mode
Maximum sensor power
Sensor
Dynamic
Range
200 ms
Minimum sensor power
1 s
6.6 s
150 ms
Typical
Settling
Times
10 dB
10 dB
10 dB
20 dB
13.5 s
5 dB