1
Power Meter Remote Operation
92
Keysight N8262A Programming Guide
Checking for Limit Failures
There are two ways to check for limit failures:
1
Use the
SENSe:LIMit:FAIL?
and
SENSe:LIMit:FCOunt?
commands for
channel limits or the
CALCulate[1|2]:LIMit:FAIL?
and the
CALCulate[1|2]:LIMit:FCOunt?
for window limits
2
Use the
STATus
command subsystem
Using SENSe and CALCulate
Using
SENSe
to check the channel limit failures in
would return the
following results:
Use the equivalent
CALCulate
commands for checking window limit failures.
Table 1-3
Range of values for window limits
Window
Units
Default
Maximum
Minimum
Maximum
Minimum
dB
+200 dB
–
180 dB
60 dB
–
120 dB
dBm
+230 dBm
–
150 dBm
90 dBm
–
90 dBm
%
999.9 X%
100.0 a%
100.0 M%
100.0 p%
W
100.000 XW
1.000 aW
1.000 MW
1.000 pW
SENSe:LIMit:FAIL?
Returns 1 if there has been 1 or more
limit failures or 0 if there have been
no limit failures. In this case 1 is
returned.
SENSe:LIMit:FCOunt?
Returns the total number of limit
failures, in this case 2.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...