TRIGger Subsystem
14
Keysight N8262A Programming Guide
571
Example
Reset Condition
On reset (
*RST
), this command is set to
OFF
.
On preset (
SYSTem:PRESet
) and instrument power-up, when entering local mode,
if
TRIGger[:SEQuence[1]|2]:SOURce
is set to
INT[[1]|2]
or
EXT
,
INITiate:CONTinuous
is not changed. For other trigger sources,
INITiate:CONTinuous
is set to
ON
.
Query
INITiate[1]|2:CONTinuous?
The query enters a 1 or 0 into the output buffer.
– 1 is returned when there is continuous triggering
– 0 is returned when there is only a single trigger
Query Example
INIT2:CONT ON
This command places Channel B in the wait
for trigger state.
INIT2:CONT?
This command queries whether Channel B is
set for single or continuous triggering.