STATus Subsystem
11
Keysight N8262A Programming Guide
431
STATus:OPERation:CALibrating[:SUMMary]
The operation status calibrating summary register set contains information on the
calibrating status of the power meter.
The following bits in these registers are used by the power meter:
These bits are set at the beginning of zeroing (
CALibration:ZERO:AUTO ONCE
) and
at the beginning of calibration (
CALibration:AUTO ONCE
). Also for the compound
command/query
CALibration[:ALL]?
, this bit is set at the beginning of the
calibration sequence.
These bits are cleared at the end of zeroing or calibration.
Syntax
Bit Number
Decimal Weight
Definition
0
-
Not used
1
2
Channel A CALibrating Status
2
4
Channel B CALibrating Status
3-15
-
Not used
STAT
:OPER
:CAL
:SUMM