10
SENSe Subsystem
384
Keysight N8262A Programming Guide
Parameters
Example
Reset Condition
On reset,
GAIN2
is set to 0.00 dB.
Query
[SENSe[1]]|SENSe2:CORRection:GAIN2[:INPut][:MAGNitude]? [MIN|MAX]
The query returns the current setting of the channel offset or the values
associated with
MIN
and
MAX
.
Item
Description/Default
Range of Values
numeric_value
A numeric value:
–
DEF
: the default is 0.00 dB
–
MIN
: –100 dB
–
MAX
: +100 dB
–100 to +100 dB
DEF
MIN
MAX
CORR:GAIN2 50
This command sets a channel offset of 50 dB
for Channel A.