MEASurement Commands
132
Keysight N8262A Programming Guide
CONFigure [1] |2|3|4 Commands
The
CONFigure
commands are used on the specified window/measurement to set:
– The expected power level being measured.
– The resolution of the window/measurement.
– The channel(s) on which the measurement is to be made.
The
CONFigure
commands do not make the power measurement after setting the
configuration. Use
READ?
, or alternatively use
INITiate
followed by a
FETCh?
to
make the measurement.
The
CONFigure
command also applies the following defaults to the channel(s)
which are in the specified window (the channel(s) in the window are specified in
the
<source list>
parameter):
Default Settings
Description
INITiate:CONTinuous OFF
Sets the power meter to make one trigger cycle when INITiate
is sent.
TRIGger:SOURce IMMediate
When
TRIG:SOUR
is set to
BUS
or
HOLD
, sets the power meter
to make the measurement immediately a trigger is received.
TRIGger:DELay:AUTO ON
Enables automatic delay before making the measurement.
SENSE:AVERage:COUNt:AUTO ON
Enables automatic filter length selection.
SENSE:AVERage:STATe ON
Enables averaging.
Summary of Contents for N8262A
Page 1: ...Keysight N8262A P Series Modular Power Meter Programming Guide ...
Page 6: ...6 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 48: ...48 Keysight N8262A Programming Guide ...
Page 50: ...50 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 54: ...54 Keysight N8262A Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 648: ...16 SERVice Subsystem 648 Keysight N8262A Programming Guide ...
Page 649: ...SERVice Subsystem 16 Keysight N8262A Programming Guide 649 ...
Page 668: ...17 IEEE 488 2 Command Reference 668 Keysight N8262A Programming Guide Syntax STB ...