2
General Power Meter Functions
70
Keysight N8262A User’s Guide
Procedure
– Click .
The
Channel Gate Setup
screen is displayed.
Figure 2-23
Gate Setup screen
1
Highlight the
Gate Start
you want to configure using the
,
,
and
softkeys.
NOTE
shows measurement results using this example of the gates
placement.
NOTE
This procedure uses the
Gates Setup
under the
Channel Setup
menu.
Alternatively, you can use a more visual method, when in the Graphical Trace
Mode (Gate Control Menu), to set the channel’s Gate Control. (See
Gates Setup
NOTE
The gate start time is relative to the trigger event. Positive values set a
measurement gate, to a maximum time of 1 second, after the trigger. Negative
time gate start values set a measurement gate, to a maximum time of 1 second,
before the trigger.