Keysight N778xC Series Polarization Instruments Programming Guide
77
Polarization Commands
5
Polarization Synchronous Scrambler Commands
Command:
:PCONtroller:STOP
Syntax:
:PCONtroller:STOP
Description:
This command stops the Polcontroller.
Parameters:
None
Response:
None
Example:
:PCONtroller:STOP
Affects
N7785C / N7786C / N7788C
Command:
:PCONtroller:WPLAtes
Syntax:
:PCONtroller:WPLAtes<wsp><orientation waveplate 1>,<retardation waveplate1>,...
Description:
Sets the orientation and retardation of the polarization controller waveplates.
The data consists of 12 double values. The order is given by:
Orientation Waveplate 1, Retardation Waveplate 1,
Orientation Waveplate 2, Retardation Waveplate 2,
Orientation Waveplate 3, Retardation Waveplate 3,
Orientation Waveplate 4, Retardation Waveplate 4,
Orientation Waveplate 5, Retardation Waveplate 5,
Orientation Waveplate 6, Retardation Waveplate 6
The orientation is given in rad, the retardation is given in fractions of Lambda. A retardation of 0.25 corresponds to a quarter
waveplate. The maximum retardation value is 0.25.
Parameters:
Double array with 12 values
Response:
None
Example:
:PCONtroller:WPLAtes 0,0.25,0,0.25,0,0.25,0,0.25,0,0.25,0,0.25
Affects
N7785C / N7786C / N7788C
Command:
:PCONtroller:WPLAtes?
Syntax:
:PCONtroller:WPLAtes?
Description:
Returns the orientation and retardation of the polarization controller waveplates
Parameters:
None