Programming the Output
Set the Output Voltageand Low Voltage Limit
Set the Output Turn-On/Turn-Off Mode
SHOCK HAZARD, LETHAL VOLTAGES - Some models generate output voltages up
to 2,000 VDC! Ensure that all instrument connections, load wiring, load and sense
connections are insulated using high-voltage wire. Always attach the safety cover
provided, so that no accidental contact with lethal voltages can occur.
When the unit is first turned on, it may take about 30 seconds or so to initialize the
instrument before it is ready for use.
Set the Output Priority Mode
Select either voltage priority or current priority operating mode. Refer to
for
more information.
Voltage priority - this mode keeps the output voltage constant. The output voltage remains at its
programmed setting, provided the load current remains within the + or - current limit setting.
Current priority - this mode keeps the output current constant. The output current remains at its
programmed setting, provided the load voltage remains within the voltage limit setting.
Front Panel Menu Reference
SCPI Command
Select
Output\Mode
.
Select Voltage or Current priority. Then press
Select
.
To specify current or voltage priority mode:
When switching between modes, the output is turned off and the output settings
revert to their power-on or RST values.