Front Panel Menu Reference
SCPI Command
Select Transient\TrigSource.
To select immediate triggers, select Imm.
To select Bus triggers, select Bus.
To select digital pin 5 as the trigger, select Pin 5, or EXT.
To select expression1 as the trigger, select Expr1.
To select Bus triggers:
To select digital pin 5 as the trigger:
To select expression1 as the trigger:
Initiate the transient system
When the unit is turned on, the trigger system is in the idle state. In this state, the trigger system is
disabled, ignoring all triggers. The INITiate commands enable the trigger system to receive triggers.
Front Panel Menu Reference
SCPI Command
Select Transient\Control.
Scroll to Initiate. Then press Select.
To initiate the transient trigger system:
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving the
INITiate:TRANsient command. If a trigger occurs before the trigger system is ready for it, the trigger
will be ignored. You can test the WTG_tran bit in the operation status register to know when the
instrument is ready to receive a trigger after being initiated.
Front Panel Menu Reference
SCPI Command
Select Transient\Control.
The Trig state field indicates "Initiated".
To query the WTG_tran bit (bit 4):
If a bit value of 16 is returned in the query, the WTG_tran bit is true, and the instrument is ready to
receive the trigger signal. Refer to
Unless INITiate:CONTinuous:TRANsient is programmed, the instrument executes
one transient each time a trigger signal is received. Thus, it will be necessary to
initiate the trigger system each time another triggered transient is desired.
Trigger the transient
The trigger system is waiting for a trigger signal in the initiated state. You can immediately trigger the
transient as follows:
Front Panel Menu Reference
SCPI Command
Select Transient\Control.
Select Trigger to generate an immediate trigger signal
regardless of the trigger source setting.
To generate a transient trigger:
Alternatively, if the trigger source is BUS, you can also
program a
or an IEEE-488 <get> command.
Keysight N6900/N7900 Series Operating and Service Guide
119
4 Using the Advanced Power System
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...