![Keysight Technologies N6900 Series Operating And Service Manual Download Page 265](http://html1.mh-extra.com/html/keysight-technologies/n6900-series/n6900-series_operating-and-service-manual_1962217265.webp)
STATus:OPERation:CONDition?
Queries the
for the
group. This is a read-only register, which holds
the live (unlatched) operational status of the instrument. Reading the Operation Status Condition
register does not clear it.
Parameter
Typical Return
(none)
<bit value>
Read the operation status condition register: STAT:OPER:COND?
l
The value returned is the binary-weighted sum of all bits set in the register. For example, with bit
3 (value 8) and bit 5 (value 32) set (and corresponding bits enabled), the query r40.
l
The condition register bits reflect the current condition. If a condition goes away, the cor-
responding bit is cleared.
l
*RST clears this register, other than those bits where the condition still exists after *RST.
STATus:OPERation:ENABle <
value
>
STATus:OPERation:ENABle?
Sets the value of the
for the
group. The enable register is a mask for
enabling specific bits from the Operation Event register to set the OPER (operation summary) bit of the
Status Byte register.STATus:PRESet clears all bits in the enable register.
Parameter
Typical Return
A decimal value corresponding to the binary-weighted sum of
the register's bits.
<bit value>
Enable bit 3 and 4 in the enable register: STAT:OPER:ENAB 24
l
For example, with bit 3 (value 8) and bit 5 (value 32) set (and corresponding bits enabled), the
query r40.
l
*CLS does not clear the enable register, but does clear the
.
STATus:OPERation:NTRansition <
value
>
STATus:OPERation:NTRansition?
STATus:OPERation:PTRansition <
value
>
STATus:OPERation:PTRansition?
Sets and queries the value of the
(Negative-Transition) and
(Positive-Transition) registers.
These registers serve as a polarity filter between the Operation Condition and Operation Event
registers.
When a bit in the NTR register is set to 1, then a 1-to-0 transition of the corresponding bit in the
Operation Condition register causes that bit in the Operation Event register to be set.
6 SCPI Programming Reference
265
Keysight N6900/N7900 Series Operating and Service Guide
Summary of Contents for N6900 Series
Page 1: ...Keysight Advanced Power System N6900 N7900 Series Operating and Service Guide...
Page 2: ......
Page 63: ...1 kW Models Keysight N6900 N7900 Series Operating and Service Guide 63 1 Quick Reference...
Page 64: ...2 kW Models 1 Quick Reference 64 Keysight N6900 N7900 Series Operating and Service Guide...
Page 78: ...2 Installing the Instrument 78 Keysight N6900 N7900 Series Operating and Service Guide...