[SOURce:]ARB:SEQuence:STEP:CURRent:RAMP:END:TIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:RAMP:END:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:RAMP:END:TIMe<
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:RAMP:END:TIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the time in seconds, after the ramp completes, that the ending current or voltage level
persists for Arb sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 0
<end time>
Step 0 - 99, *RST 0
(none)
Programs an end time for the current ramp: ARB:CURR:RAMP:END:TIM 2,(@1)
Programs an end time for the voltage ramp:ARB:VOLT:RAMP:END:TIM 2,(@1)
Programs an end time for step 0: ARB:SEQ:STEP:CURR:RAMP:END:TIM 2, 0,(@1)
[SOURce:]ARB:CURRent:RAMP:RTIMe <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:RAMP:RTIMe? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:RTIMe<
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:RTIMe? [MIN|MAX,] (@<chanlist>)
Specifies the rise time of the ramp in seconds. Referenced to t1 in the Ramp diagrams.
[SOURce:]ARB:SEQuence:STEP:CURRent:RAMP:RTIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:RAMP:RTIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:RAMP:RTIMe <
value
>, <
step#
>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:RAMP:RTIMe? [MIN|MAX,] <
step#
>, (@<chanlist>)
Specifies the rise time of the ramp in seconds for Arb sequence steps.
Parameter
Typical Return
0 – 262.144 | MIN | MAX, *RST 1
<rise time>
Step 0 - 99, *RST 0
(none)
Programs a rise time for the current ramp: ARB:CURR:RAMP:RTIM 10,(@1)
Programs a rise time for the voltage ramp: ARB:VOLT:RAMP:RTIM 10,(@1)
Programs a rise time for step 0: ARB:SEQ:STEP:CURR:RAMP:RTIM 10, 0,(@1)
[SOURce:]ARB:CURRent:RAMP:STARt[:LEVel] <
value
>, (@<chanlist>)
[SOURce:]ARB:CURRent:RAMP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:STARt[:LEVel] <
value
>, (@<chanlist>)
[SOURce:]ARB:VOLTage:RAMP:STARt[:LEVel]? [MIN|MAX,] (@<chanlist>)
Specifies the initial current or voltage before the ramp occurs. Referenced to I0 and V0 in the Ramp
diagrams.
Keysight N6705C Operating and Service Guide
245
7 SCPI Programming Reference
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......