Constant-Dwell
[SOURce:]ARB:CURRent:CDWell[:LEVel] <
value
>{<,
value
>},(@<chanlist>)
[SOURce:]ARB:CURRent:CDWell[:LEVel] ? (@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell[:LEVel] <
value
>{<,
value
>},(@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell[:LEVel] ? (@<chanlist>)
Specifies the level of each current or voltage point in a constant-dwell Arb.
Current and voltage CD ARBs share settings, so setting the current CD ARB resets the voltage CD
ARB level to its default value and vice-versa. A comma-delimited list of up to 65,535 points may be
programmed.
Parameter
Typical Return
Current: 0 - maximum
Voltage: 0 - maximum
The maximum value varies, depending on the current or voltage rating
of the power module. *RST 1 point set to the minimum value.
<value> [,<value>]
or
<block> [,<block>]
Program five current points: ARB:CURR:CDW 5,4,3,2,1,(@1)
Program five voltage points: ARB:VOLT:CDW 20,21,22,23,24,(@1)
l
For better performance, the list of voltage points can be sent as a definite length binary block
instead of an ASCII list.
l
The return format depends on the settings of the FORMat:BORDer and FORMat[:DATA] com-
mands. When the data format is set to ASCII, returned values are comma separated. Only one
channel can be queried when the format is ASCII. When the data format is set to REAL, data is
returned as single precision floating point numbers in definite length arbitrary block response
format. Each block contains all the records for one of the channels given in the <chanlist> para-
meter. If multiple channels are given, then each definite length arbitrary block of data is separated
by a comma.
[SOURce:]ARB:CURRent:CDWell:DWELl <
value
>,(@<chanlist>)
[SOURce:]ARB:CURRent:CDWell:DWELl? (@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell:DWELl <
value
>,(@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell:DWELl? (@<chanlist>)
Specifies the dwell time in seconds for each current or voltage point in a constant-dwell ARB
Current and voltage CD ARBs share settings, so setting this parameter for a current CD ARB also
changes the voltage dwell value and vice versa. Programmed values can range from 10.24
microseconds to 0.30 seconds. Values are rounded to the nearest 10.24-microsecond increment.
Parameter
Typical Return
0.00001024 - 0.30, *RST 0.001
<dwell value>
Program a constant-dwell current time: ARB:CURR:CDW:DWEL 0.2,(@1)
Program a constant-dwell voltage time: ARB:VOLT:CDW:DWEL 0.2,(@1)
7 SCPI Programming Reference
236
Keysight N6705C Operating and Service Guide
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......