To select the trigger source for step 2:
To end the sequence at the last Arb value:
To make the entire sequence repeat twice:
To set up the transient trigger system and trigger the sequence:
VOLT:MODE ARB,(@1)
TRIG:ARB:SOUR BUS
OUTP ON,(@1)
INIT:TRAN (@1)
*TRG
Arb Trigger Sources
Specify a Trigger source for the arbitrary waveforms. The same trigger source will be used to trigger all
of the arbitrary waveforms
From the front panel: Press the Arb key, then select the Trigger Source field.
Trigger Source
Description
Arb Run/Stop Key
The front panel Run/Stop key.
BNC Trigger In
The rear trigger input BNC connector
Remote Command
A remote interface command.
From the remote interface:
Select from one of the following SCPI trigger sources:
Keysight N6705C Operating and Service Guide
105
3 Using the Source Functions
Summary of Contents for N6705C
Page 76: ......
Page 353: ...Status Diagram Keysight N6705C Operating and Service Guide 353 7 SCPI Programming Reference ...
Page 376: ...8 Verification and Calibration 376 Keysight N6705C Operating and Service Guide ...
Page 389: ...Keysight N6705C Operating and Service Guide 389 8 Verification and Calibration ...
Page 434: ......
Page 438: ...9 Service and Maintenance 438 Keysight N6705C Operating and Service Guide ...
Page 460: ...9 Service and Maintenance 460 Keysight N6705C Operating and Service Guide ...
Page 461: ...Keysight N6705C Operating and Service Guide 461 9 Service and Maintenance ...
Page 462: ...9 Service and Maintenance 462 Keysight N6705C Operating and Service Guide ...
Page 463: ...Keysight N6705C Operating and Service Guide 463 9 Service and Maintenance ...
Page 470: ...Mezzanine Parts 9 Service and Maintenance 470 Keysight N6705C Operating and Service Guide ...
Page 471: ...Front Panel Parts Keysight N6705C Operating and Service Guide 471 9 Service and Maintenance ...
Page 476: ......