2 Installation
28
Keysight N6700 User’s Guide
Step 3. Use a T10 Torx driver and install the screws at each end of the power
module. Because the RFI strips apply upward pressure, continue pushing
down on the module until the screws are tight.
Step 4. Replace the blower cover when finished. Carefully fit the spring clips
under the lip of the power modules.
Rack Installation
CAUTION
You cannot use support rails for rack mounting your instrument.
Support rails would block the airflow needed for cooling.
Use Rack Mount kit (Option 908) to rack mount your instrument.
The Rack Mount Kit is also available by ordering part number N6709A.
Keysight N6700 MPS mainframes can be mounted in a 19-inch EIA rack
cabinet. They are designed to fit in one rack-unit (1U) of space. Do not
block the air intake and exhaust at the sides of the unit, or the exhaust at
the rear of the unit.
Tools required: Phillips driver, T22 Torx driver, T10 Torx driver
Step 1. Install eight clip-nuts on the rack frame (2 in each corner) where your
instrument will be located.