Sys Burst Tests
8
Test 2.3.5-SYS-RX Differential Termination Disable Time (T
TERM-OFF-SYS-RX
)
It is available for all hardware configurations.
Figure 162
Test 2.3.5-SYS-RX Differential Termination Disable Time
(T
TERM-OFF-SYS-RX
)
Purpose
It determines that the SYS-RX disables the termination within the required
time.
Procedure
The data generator sends Burst-mode SYS CJTPAT signaling with nominal
amplitude settings while the DUT is configured for Terminated-mode.
Then, the T
TERM-OFF-SYS-RX
(PWM-RX differential termination disable time)