Sys Burst Tests
8
490mV and V
CM-RX
= 330mV for NT); minimum differential voltage and
minimum common-mode (V
DIF-RX
= 60mV and V
CM-RX
= 25mV for RT,
V
DIF-RX
= 120mV and V
CM-RX
= 25mV for NT). At each step, the BER is
measured and compared with the Target BER
to consider the pass/fail
result.
In
Expert Mode
, the common-mode voltage component of each Level Pair
is tested over the given Offset Range. In this case, the result will show the
minimum and maximum offset values at which the DUT works properly and
the test will pass if those values are within the spec limits.
This test is performed at each selected lane.
Connection Diagram
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test
test.
Parameters
The parameters used in
Expert Mode
for this test, are as follows:
• CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Peaking (if CDR is selected in the Special Parameter dialog, see
• CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
)
• CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
• Target BER
• BerReader Init Mode
• Test Sequence
• Re-Init sequence after Reset DUT
• Lane Under Test Termination Model
• Level Pairs
• Offset Range
These parameters are listed in alphabetical order in
Dependencies
The calibrations required for this test are as follows:
Summary of Contents for N5990A
Page 1: ...User Guide Keysight N5990A MIPI M PHY Receiver Test...
Page 62: ......
Page 86: ...86 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 88: ...88 Keysight N5990A MIPI M PHY Receiver Test User Guide 4 Calibrations...
Page 214: ......
Page 234: ......
Page 268: ......
Page 298: ......