Keysight N5990A MIPI M-PHY Receiver Test User Guide
137
Calibrations
4
Figure 77
Example HTML Viewer for Accumulated Voltage Calibration Terminated /
Into Open
Table 17
Accumulated Voltage Calibration Terminated / Into Open data table
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If the set-up is wrong, the calibration procedure fails due to the non-monotonic values
measured.
Set Differential Amplitude
It is the set value for differential amplitude voltage.
Measured Eye Opening
It is the measured value of eye opening according to the set differential voltage.
Maximum V_diff_ac_tx
It is the allowed maximum accumulated voltage value according to the specification.