Keysight N5990A MIPI M-PHY Receiver Test User Guide
121
Calibrations
4
High Frequency SJ Calibration
This procedure is available for all hardware configurations.
Figure 69
High Frequency SJ Calibration
Purpose
This procedure calibrates the short term sinusoidal jitter that is added to
the generated signal.
This calibration should be run once at each selected HS data rate. For
spec 4.1, this calibration is only mandatory for B-series (for A-series, the
procedure is shown in the test tree as informative).