Keysight N5990A MIPI M-PHY Receiver Test User Guide
103
Calibrations
4
Results
An example of a HTML Viewer for the ISI Calibration procedure is shown in
. The results comprise of the following:
• A calibration data graph.
• The common parameter list (in
Expert Mode
).
• A data table for the ISI that is being calibrated (Refer to
).
Table 9
ISI Calibration data table
Parameter Name
Parameter Description
Result
"Pass"/"Fail" - If it is "Fail", the measured or obtained values are not monotonic.
Data Rate
HS data rate at which the ISI was calibrated.
Measured Offset
It is the measured value of ISI.