7-78 Keysight N5247B Service Guide
Repair and Replacement Procedures
Removing and Replacing the A70 or A75 Low Frequency Extension (LFE) Board
7-
11.
Disconnect the ribbon cable (item
) from the A70 LFE board. Refer to
Figure 7-41
A70 LFE Board Removal (N5291-60001, 0515-0665 (x2)). (A75 LFE Board
(N5291-60005) Removal is Similar.)
12.
Remove the two screws from the LFE board (item
). Save for reuse.
Refer to
.
13.
Lift the LFE board (N5291-60001) and remove it from the analyzer. Refer
.
Replacement Procedure
1.
Reverse the order of the removal procedure.
Attach the new A70/A75 LFE board to the standoffs using the screws
removed from the old one.
2.
Perform the post-repair adjustments, verifications, and performance tests
that pertain to this removal procedure. Refer to
.