Keysight N5224A/25A Service Guide 8-7
General Purpose Maintenance Procedures
Error Terms
—
“Front Panel Access to Error Terms” on page 8-7
Programmatically
—
“GPIB Access to Error Terms” on page 8-7
—
“COM/DCOM Access to Error Terms” on page 8-7
Manual Access to Error Terms
Front Panel Access to Error Terms
To access the error terms from the front panel, perform the following steps:
1.
Press RESPONSE
, then
. Verify that
is ON. If not, press the softkey to toggle it ON.
The
Cal Set Viewer
toolbar appears directly above the trace window.
2.
In the
Cal Set
list, select the desired cal set.
3.
Click the
Standards or Error Terms
button to view the raw measurement
data from the standard or the corrected error term data.
4.
In the
Standard
or
Error Terms
list, select the standard or error terms to
view. Click the Enable check box to enable the selection.
5.
Compare the displayed measurement trace to previously measured data
or to the uncorrected performance specifications listed in
.
6.
Print numerical data or print a plot of the measurement results.
Programmatic Access to Error Terms
GPIB Access to Error Terms
You can access error terms by way of GPIB with Standard Commands for
Programmable Instruments (SCPI).
For more information on GPIB and SCPI, refer to the embedded help in the
analyzer. Type in keyword “errors, systematic” in the index.
COM/DCOM Access to Error Terms
You can access error terms by way of Component Object Model (COM) or
Distributed Component Object Model (DCOM) software architecture.
For more information on COM and DCOM, refer to the embedded help in the
analyzer. Type in keyword “errors, systematic” in the index.
Ensure that calibration correction is active by pressing RESPONSE
and verifying that the softkey label reads
Correction
ON/off
. If not, press
the
key and it will toggle to read
.
Cal
Correction on/OFF
Correction ON/off
Cal
Manage Cals
Cal Set Viewer ON/off