7-14 Keysight N5224A/25A Service Guide
Repair and Replacement Procedures
Removing and Replacing the A4–A17 Boards
7-
Removing and Replacing the A4–A17 Boards
Tools Required
— T-10 TORX driver (torque to 9 in-lb or 1.02 N.m)
— T-20 TORX driver (torque to 21 in-lb or 2.38 N.m)
— 5/16 inch open-end torque wrench (torque to 10 in-lb or 1.13 N.m)
— 9 mm socket or open-end wrench (torque to 21 in-lb or 2.38 N.m)
— ESD grounding wrist strap
Removal Procedure
Refer to
for this procedure.
1.
Disconnect the power cord.
2.
Remove the outer and inner covers. Refer to
.
3.
A5 and A10 source boards:
a.
A10 source board only: on the top side of the analyzer, disconnect
all visible semirigid cables from the A10, A12, and A13 boards (item
①
).
b.
A5 source board only: on the top side of the analyzer, disconnect all
visible semirigid cables from the A4, A7, and A8 boards.
c.
On the top side of the analyzer, remove two screws (item
➁
), one at
each end of the board, from the source board to be removed.
d.
Lift the two extractors (item
③
), one at each end of the board.
Adjust the slack in the gray flexible cable to move it out of the way,
and lift the board out of the chassis.
4.
A7, A8, A12, A13 boards: on the bottom side of the analyzer, disconnect
cables (item
①
) from the cable brackets.
5.
A4, A7, A8, A12, A13, A14, A15, A16, and A17 boards:
a.
Remove all cables connected to the top of the board to be removed.
Note the location of each cable for reinstallation.
b.
Remove two screws (item
➁
), one at each end of the board, from
the board to be removed (Exception: A14 does not have screws).
c.
Lift the two extractors (item
③
), one at each end of the board, and
lift the board.
d.
Before removing the board, check the bottom of the board for any
attached cables.
Replacement Procedure
1.
Reverse the order of the removal procedure.
Any cables that are removed should be labeled for reinstallation later.