312
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Digital Signal Interface Module (Option 003/004)
Operating the N5102A Module in Input Mode
2.
Select the logic type required for the device being tested.
A caution message is displayed whenever a change is made to the logic types, and a softkey
selection appears asking for confirmation.
3.
Refer to
. Press the
Port Config
softkey.
In this menu, select either a serial, parallel, or parallel interleaved data transmission.
4.
Select the port configuration for the device being tested.
Configuring the Clock Signal
1.
Refer to
. Press the
Clock Setup
softkey.
Figure 10-19
Clock Setup Menu Location
From this softkey menu, set all of the clock parameters that synchronize the data between the
N5102A module and the device. From this menu, the clock signal phase can be changed so the
clock occurs during the valid portion of the data.
shows the clock setup menu.
If the device or external clock does not match the frequency, one of the following error
messages will appear on the signal generator:
803
Digital module input FIFO overflow error; There are more
samples being produced than can be consumed at the
current clock rate. Verify that the digital module clock
is set up properly.
This error is reported when the digital module clock setup is not
synchronized with the rate the samples are entering the digital
module. Verify that the input clock rate matches the specified
Within the data and clock setup softkey menus, some softkeys function
relative to the current configuration. Softkeys that are grayed out are not
available for the current setup. Refer to the help text to determine which
parameter is causing the softkey to be unavailable. Press the
Help
hardkey
on the signal generator front panel and then the softkey that is
unavailable.
Accesses the Clock Setup
Menu