25
Second Harmonic Distortion (SHD)
Procedure
Procedure
1.
Configure the equipment as shown in
.
2.
Press
Mode
,
Spectrum Analyzer
,
Mode Preset
on the signal analyzer and
Preset the synthesized sweeper.
3.
Set up the synthesized sweeper by pressing:
Frequency
, 40,
MHz
Amplitude
, –10,
dBm
4.
Set up the signal analyzer by pressing:
Input/Output
,
More
,
Freq Ref In
,
External
FREQ Channel
,
Center Freq
, 40,
MHz
SPAN X Scale
,
Span
, 1,
MHz
5.
On the analyzer, press
Peak Search
.
6.
Adjust the synthesized sweeper amplitude for a signal analyzer display of
–10 dBm
±
0.1 dB.
7.
On the analyzer, activate the marker delta function by pressing the
Marker
and
Delta
keys.
8.
On the signal analyzer, press:
FREQ Channel
,
Center Freq
, 80,
MHz
Meas Setup
,
Average/Hold Number
, 20,
Enter
Trace/Detector, Trace Average
Single
9.
Press Peak Search. Enter the displayed value under the Measured Second
Harmonic Distortion (dBc) heading in
Table 4-1
Second Harmonic Distortion Results
Applied Frequency
(MHz)
Measured Second Harmonic Distortion
(dBc)
Mixer Level
(dBm)
Specification
(dBc)
40
–10
–55