![Keysight Technologies M9260A Startup Manual Download Page 40](http://html1.mh-extra.com/html/keysight-technologies/m9260a/m9260a_startup-manual_1962002040.webp)
1
Inspection, Installation, and Verification
40
Keysight M9260A Startup Guide
6
Compare the measured values to the accuracy values as stipulated in
.
7
Repeat
Step 3
to
Step 6
using different output amplitude and input range
values for all voltage ranges and channels. Refer to
suggested output amplitude and input range values.
Table 1-3
Operational verification table
Output Amplitude
Input Range
Measurement
Low Limit
High Limit
0.316 V
0.32 V
Vac
0.2123 V
0.2346 V
Freq
999.998 Hz
1000.02 Hz
THD+N
< -103 dB
THD
< -110 dB
1 V
1 V
Vac
0.7000 V
0.7142 V
Freq
999.998 Hz
1000.02 Hz
THD+N
< -103 dB
THD
< -110 dB
3.16 V
3.2 V
Vac
2.2121 V
2.2568 V
Freq
999.998 Hz
1000.02 Hz
THD+N
< -103 dB
THD
< -110 dB
10 V
10 V
Vac
7.0004 V
7.1418 V
Freq
999.998 Hz
1000.02 Hz
THD+N
<
-103dB
THD
< -110 dB
32 V
Vac
7.0004 V
7.1418 V
Freq
999.998 Hz
1000.02 Hz
THD+N
< -95 dB
THD
< -105 dB
46 V
Vac
7.0004 V
7.1418 V
Summary of Contents for M9260A
Page 1: ...Keysight M9260A PXIe Audio Analyzer Startup Guide...
Page 10: ...10 Keysight M9260A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 12: ...12 Keysight M9260A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 14: ...14 Keysight M9260A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 16: ...16 Keysight M9260A Startup Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...