26
6000L Series Oscilloscopes Service Guide
2
Testing Performance
1
Turn on the test equipment and the oscilloscope. Let them warm up for 30
minutes before starting the test.
2
Set up the oscilloscope calibrator.
a
Set the oscilloscope calibrator to provide a DC offset voltage at the
Channel 1 output.
b
Use the multimeter to monitor the oscilloscope calibrator DC output
voltage.
3
Use the 8-by-2 test connector and the BNC cable assembly to connect digital
channels D0-D7 to one side of the BNC Tee. Then connect the D0-D7 ground
lead to the ground side of the 8-by-2 connector. See
.
Figure 2
Setting Up Equipment and Test Connector for the Threshold Test
Test
Connector
BNC-Banana
cable
Oscilloscope
Calibrator
Digital
Multimeter
BNC Tee
Probe
Cables
Channels
8 - 15
Channels
0 - 7
6000L Series Oscilloscope