802
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
27
:SBUS<n> Commands
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC
Command Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC <dlc>
<dlc> ::= integer between -1 (don't care) and 15, in NR1 format,
when trigger is in DATA mode.
<dlc> ::= integer between -1 (don't care) and 255, in NR1 format,
when trigger is in LDATa mode.
For the trigger types that let you trigger on data, the
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC command specifies the data length
code of the CXPI frame you wish to trigger on.
This will also affect the number of data bytes you can specify in the trigger.
Query Syntax
:SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC?
The :SBUS<n>:CXPI:TRIGger:PATTern:INFO:DLC? query returns the DLC trigger
value setting.
Return Format
<dlc><NL>
<dlc> ::= integer between -1 (don't care) and 15, in NR1 format,
when trigger is in DATA mode.
<dlc> ::= integer between -1 (don't care) and 255, in NR1 format,
when trigger is in LDATa mode.
See Also
•
•
":SBUS<n>:CXPI:TRIGger:IDFilter"
•
•
":SBUS<n>:CXPI:TRIGger:PATTern:DATA"
•
":SBUS<n>:CXPI:TRIGger:PATTern:DATA:LENGth"
•
":SBUS<n>:CXPI:TRIGger:PATTern:DATA:STARt"
•
":SBUS<n>:CXPI:TRIGger:PATTern:ID"
•
":SBUS<n>:CXPI:TRIGger:PATTern:INFO:CT"
•