740
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
27
:SBUS<n> Commands
:SBUS<n>:A429:AUTosetup
Command Syntax
:SBUS<n>:A429:AUTosetup
The :SBUS<n>:A429:AUTosetup command automatically sets these options for
decoding and triggering on ARINC 429 signals:
• High Trigger Threshold: 3.0 V.
• Low Trigger Threshold: -3.0 V.
• Noise Reject: Off.
• Probe Attenuation: 10.0.
• Vertical Scale: 4 V/div.
• Serial Decode: On.
• Base (:SBUS<n>:A429:BASE): HEX.
• Word Format (:SBUS<n>:A429:FORMat): LDSDi (Label/SDI/Data/SSM).
• Trigger: the specified serial bus (n of SBUS<n>).
• Trigger Mode (:SBUS<n>:A429:TRIGger:TYPE): WSTArt.
Errors
•
See Also
•
•
•
•
"Introduction to :SBUS<n> Commands"
•
•