340
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
13
:DISPlay Commands
:DISPlay:PERSistence
Command Syntax
:DISPlay:PERSistence <value>
<value> ::= {MINimum | INFinite | <time> | ADAPtive}
<time> ::= seconds in in NR3 format from 100E-3 to 60E0
The :DISPlay:PERSistence command specifies the persistence setting:
• MINimum — indicates zero persistence.
• INFinite — indicates infinite persistence.
• <time> — for variable persistence, that is, you can specify how long acquisitions
remain on the screen.
• ADAPtive — all previous acquisitions that have taken place in hardware since
the last controller PC screen capture are displayed.
Use the :DISPlay:CLEar command to erase points stored by persistence.
Query Syntax
:DISPlay:PERSistence?
The :DISPlay:PERSistence? query returns the specified persistence value.
Return Format
<value><NL>
<value> ::= {MIN | INF | <time> | ADAP}
See Also
•
"Introduction to :DISPlay Commands"
•