1258
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
34
:WMEMory<r> Commands
:WMEMory<r>:YOFFset
Command Syntax
:WMEMory<r>:YOFFset <offset> [<suffix>]
<r> ::= 1 to (# ref waveforms) in NR1 format
<offset> ::= vertical offset value in NR3 format
<suffix> ::= {V | mV}
The :WMEMory<r>:YOFFset command sets the value that is represented at center
screen for the selected reference waveform.
The range of legal values varies with the value set by the :WMEMory<r>:YRANge
or :WMEMory<r>:YSCale commands. If you set the offset to a value outside of the
legal range, the offset value is automatically set to the nearest legal value. Legal
values are affected by the probe attenuation setting.
Query Syntax
:WMEMory<r>:YOFFset?
The :WMEMory<r>:YOFFset? query returns the current offset value for the selected
reference waveform.
Return Format
<offset><NL>
<offset> ::= vertical offset value in NR3 format
See Also
•
, “:WMEMory<r> Commands,” starting on page 1251
•
•
•
•