1074
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
31
:TRIGger Commands
:TRIGger:LEVel:HIGH
Command Syntax
:TRIGger:LEVel:HIGH <level>, <source>
<level> ::= 0.75 x full-scale voltage from center screen in NR3 format
for internal triggers
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :TRIGger:LEVel:HIGH command sets the high trigger voltage level voltage for
the specified source.
High and low trigger levels are used with runt triggers and rise/fall time
(transition) triggers.
Query Syntax
:TRIGger:LEVel:HIGH? <source>
The :TRIGger:LEVel:HIGH? query returns the high trigger voltage level for the
specified source.
Return Format
<level><NL>
See Also
•
"Introduction to :TRIGger Commands"
•
•
•
":TRIGger:TRANsition Commands"
•