:SYSTem:PROTection:LOCK
Command Syntax
:SYSTem:PROTection:LOCK <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :SYSTem:PROTection:LOCK command disables the fifty ohm impedance
setting for all analog channels.
Query Syntax
:SYSTem:PROTection:LOCK?
The :SYSTem:PROTection:LOCK? query returns the analog channel protection lock
status.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also
•
"Introduction to :SYSTem Commands"
NOTE
Be careful when turning ON the :SYSTem:PROTection:LOCK because there is no visual
indication of this setting in the front-panel user interface (other than a disabled 50
Ω
channel