Testing Performance
2
6000 X-Series Oscilloscopes Service Guide
49
Test External Trigger Sensitivity (all models)
This test applies to all models.
Verify the external trigger sensitivity at these settings:
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the power splitter input.
b
Connect one output of the power splitter to the Aux Trig input through a 50
Ω
feedthrough termination.
c
Connect the power sensor to the other output of the power splitter.
Table 17
External Trigger Sensitivity Test Settings
Input Range
100 MHz
200 Mhz
1.6 V
40 mV
pp
70 mV
pp
8 V
200 mV
pp
350 mV
pp