356
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide
23
CAN/LIN Triggering and Serial Decode
•
Sync
(Sync Break) — The oscilloscope triggers on the rising edge at the Sync
Break exit of the LIN single-wire bus signal that marks the beginning the
message frame.
•
ID
(Frame ID) — The oscilloscope triggers when a frame with an ID equal to
the selected value is detected. Use the
Entry
knob to select the value for the
Frame ID.
•
ID & Data
(Frame ID and Data) — The oscilloscope triggers when a frame with
an ID and data equal to the selected values is detected. When triggering on
a frame ID and data:
•
To select the frame ID value, press the
Frame ID
softkey, and use the
Entry
knob.
Note that you can enter a "don't care" value for the frame ID and trigger
on data values only.
•
To set up the number of data bytes and enter their values (in hexadecimal
or binary), press the
Bits
softkey to open the LIN Bits Menu.
LIN Serial Decode
To set up the oscilloscope to capture LIN signals, see
NOTE
For details about using the LIN Bits Menu softkeys, press and hold the softkey in question to
display the built-in help.
NOTE
Summary of Contents for InfiniiVision 3000 X Series
Page 1: ...Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 8: ...8 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 24: ...24 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide...
Page 48: ...48 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 1 Getting Started...
Page 62: ...62 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 72: ...72 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 122: ...122 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 6 Digital Channels...
Page 128: ...128 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 7 Serial Decode...
Page 134: ...134 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 8 Display Settings...
Page 182: ...182 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 11 Trigger Mode Coupling...
Page 200: ...200 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 12 Acquisition Control...
Page 210: ...210 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 13 Cursors...
Page 252: ...252 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 15 Mask Testing...
Page 256: ...256 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 16 Digital Voltmeter...
Page 290: ...290 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 19 Print Screens...
Page 324: ...324 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 21 Web Interface...
Page 344: ...344 Keysight InfiniiVision 3000 X Series Oscilloscopes User s Guide 22 Reference...