Troubleshooting
4
90000 A-Series Service Guide
75
LED Troubleshooting
If you see a failure with the on/off switch backlight LED, replace the On/Off board.
If the LED will still not illuminate, replace the motherboard and see if this fixes the
problem. If the LED still does not work, check the on/off cable connecting the
On/Off board to the motherboard.
If any of the other LEDs are not working on the front panel, the first step is to
trouble shoot the acquisition assemblies to make sure they are working. If they are
working, replace the keyboard assembly and see if the LEDs will illuminate. If the
LEDs do not illuminate, the last step is to reload the oscilloscope software.
When reassembling the oscilloscope, be sure to reinstall all boards that were
replaced, but that were not causing the problem.