62
90000 A-Series Service Guide
4
Troubleshooting
Over-Voltage Fault
If the oscilloscope loses power almost immediately as described earlier, it is
probably an over-voltage fault.
Three groups of LEDs indicate over-voltage faults. One group is located near the
center of the top edge of the backplane assembly just above the blue and red
SATA cables.
If any of these LEDs are lit, the backplane assembly must be replaced.
The other two groups of LEDs that indicate over-voltage faults are located along
the edges of the two acquisition assemblies and are visible through the bottom fan
opening located on the left side of the oscilloscope, nearest the rear panel.