![Keysight Technologies EXA Installation Note Download Page 20](http://html1.mh-extra.com/html/keysight-technologies/exa/exa_installation-note_1961984020.webp)
20
Installation Note N9010-90041
Option EXM, External Mixing Upgrade Kit
Figure 14
W27 and W28 Properly Aligned
5.
Refer to Appendix Figure 15, “Torque Cable W27 onto W28,”. Use a ¼” open-end wrench to
prevent the SMA female connector on W28 from twisting. Use a 5/16” torque wrench to torque
the nut on W27 to 10 inch-pounds.
Figure 15