Channel 2 Adjustments (EDU33212A only)
Self calibration adjustment (channel 2)
Output impedance adjustment (channel 2)
AC amplitude (high-impedance) adjustment (channel 2)
-8 dB range flatness adjustment (channel 2)
-24 dB range flatness adjustment (channel 2)
Self calibration adjustment (channel 2)
1. Enter and begin the following setup.
Setup
SCPI command
54
[a]
Self-calibration. The output is disabled.
CAL:SET 54
[a] Constants are stored after completing this setup.
2. To proceed:
a. If your calibration procedures require you to verify the adjustment just made, exit the calibration menu and
perform
DC Offset Voltage Verification
. Be sure to do this for channel 2.
b. If you are making all of the adjustments and then verifying the instrument’s performance, continue with the
next procedure in this section.
Keysight EDU33210 Series Service Guide
41