
10 Keysight E8363B/C and E8364B/C Installation Note
General Information
Installation Procedure for the Upgrade Kit
1-
Step 3.Raise the Receiver Deck
Refer to
for this procedure.
1.
Place the analyzer bottom-side up on a flat surface.
2.
With a T-10 TORX driver, remove the four screws, (item
①
), that secure
the receiver deck.
3.
Pull the latch pin (item
✍
) towards the opposite side of the analyzer to
release the receiver deck.
4.
Lift the receiver deck to partially raise it, then release the latch pin (item
✍
). Lift the receiver deck to its fully raised position and ensure that the
latch pin latches in the raised position.
Figure 4
Receiver Deck, Raising