Keysight E5260/E5270 Programming Guide, Edition 4
1- 13
Programming Basics
Getting Started
To Disable Source/Measurement Channels
To disable the channels, send the CL command. The E5260/E5270 opens the output
switch of the specified channels. Opening the output switch disables the channel.
Example
OUTPUT @E5270;"CL 1"
This example disables channel 1 (the module installed in slot 1 of the E5260/
E5270). If you do not specify the channel, the CL command disables all channels.
To Control ASU
This function is available for the E5270B that is installed with the high resolution
SMU (HRSMU). Atto Sense and Switch Unit (ASU) has two inputs, SMU input for
the HRSMU and AUX input for the other instrument. And the ASU input to output
connection can be controlled by the following commands. When the instrument is
turned on, the SMU input will be connected to the ASU output. However, the SMU
output switch will be off at this time.
Table 1-2
ASU Input Output Connection Control
where,
slot
must be the slot number assigned to the slot that installs the HRSMU
connected to the ASU. See
,
, and
function and control commands of the ASU.
When the SMU side is connected to the ASU output, the source output on/off can be
controlled by the CN/CL command. And then the SAP
slot
, 1 command is used to
change the output connection to the AUX side. When the AUX side is connected,
the output of the instrument connected to the AUX input is appeared to the ASU
output immediately.
Previous Connection
Command
Subsequent Connection
SMU side, Output on/off
SAP
slot
, 1
AUX side
SMU side, Output off
CN
slot
SMU side, Output on
SAP
slot
, 0
AUX side
CN
slot
SAP
slot
, 0
CL [
slot
]
SMU side, Output off
SMU side, Output on
CL [
slot
]
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......