Keysight E5260/E5270 Programming Guide, Edition 4
1- 11
Programming Basics
Getting Started
To Set the Measurement Range
To set the measurement range, send the following command. For details, refer to
Chapter 4, “Command Reference.”
Example
The following example sets the voltage measurement ranging mode of channel 1 to
auto.
OUTPUT @E5270;"RV 1,0"
The following example sets the current measurement ranging mode of channel 1 to
auto, and sets the current measurement auto ranging operation by using the RM
command.
OUTPUT @E5270;"RI 1,0"
OUTPUT @E5270;"RM 1,3,90"
NOTE
To use 1 pA range of ASU
Set the 1 pA limited auto ranging mode or the 1 pA fixed range mode. Or enable the
1 pA range for the auto ranging mode by using the SAR command. See
NOTE
Range change at state change
Measurement range setting is stored independently for the local state and the remote
state. So the range change may be caused by the state change.
Command
Description
RI
Sets the current measurement range.
Available for measurement except for the high speed spot
measurement.
RV
Sets the voltage measurement range.
Available for measurement except for the high speed spot
measurement.
TI, TTI
Sets the current measurement channel and the measurement
range, and performs the high speed spot measurement.
TV, TTV
Sets the voltage measurement channel and the measurement
range, and performs the high speed spot measurement.
Summary of Contents for E5260 Series
Page 1: ...Keysight Technologies E5260 E5270 Series of Parametric Measurement Solutions Programming Guide...
Page 3: ...User s Guide B2900 9001x xxxx 201x Edition x...
Page 13: ...1 Programming Basics...
Page 63: ...2 Remote Mode Functions...
Page 108: ...2 46 Keysight E5260 E5270 Programming Guide Edition 4 Remote Mode Functions Initial Settings...
Page 109: ...3 Programming Examples...
Page 185: ...4 Command Reference...
Page 335: ...5 Error Messages...
Page 349: ......