14 Keysight E5080B ENA Series RF Network Analyzers
Installation
Environmental Requirements
1-
3. A packaged item supplied with the analyzer is missing.
4. A fault has been detected in the subsequent operation check of the
analyzer.
If an abnormality is detected in step 1, contact the company that
transported the analyzer as well as your nearest Keysight Technologies
sales and service office.
Environmental Requirements
Set up the E5080B where the following environmental requirements are met.
Operating Environments
Ensure that the operating environment meets the following requirements.
Ventilation Requirements
To ensure safety requirements, the specifications and measurement accuracy
of the analyzer, you must keep environmental temperature within the specified
range by providing appropriate cooling clearance around the analyzer or, for
the rack-mount type, by forcefully air-cooling inside the rack housing. For
more information on environmental temperature to satisfy the specifications
and measurement accuracy of the analyzer, see the specification in the
E5080B Online Help.
When the environmental temperature around the analyzer is kept within the
temperature range of the operating environment specification (See Section
“Operating Environments” on page 14
), the analyzer conforms to the
Temperature
0
˚
C to 40
˚
C
Temperature range at the
error-correction
23
˚
C
3
˚
C (< 1
˚
C deviation from the temperature
when performing the error-correction)
Humidity
20% to 80% at wet bulb temperature < +29
˚
C
(non-condensation)
Altitude
0 to 2,000 m (0 to 6,561 feet)
Vibration
0.21 G maximum, 5 Hz to 500 Hz
Above environmental requirements are NOT for the specifications and
measurement accuracy of the analyzer, but for the operating environment
of the analyzer.
Summary of Contents for E5080B
Page 1: ...Installation Guide Keysight E5080B ENA Series RF Network Analyzers...
Page 10: ...10 Keysight E5080B ENA Series RF Network Analyzers Caution Manuals for E5080B...
Page 12: ...12 Keysight E5080B Installation Guide Contents Procedure to execute the factory recovery 23...
Page 26: ...26 Keysight E5080B ENA Series RF Network Analyzers Troubleshooting System Recovery 2...