3. Perform calibration
To measure impedance accurately, recommend to
perform calibration. However, if you don’t have
calibration kit, you can skip step 3.
a. Select calibration kit
Press [Cal] then press Cal Kit, and Leaded 50 ohm
b. Measure calibration standard
Press [Cal] then press Calibrate
Press Impedance calibration
Make open state then press Open
Make short sate then press Short
Make load state then press Load
c. Activate calibration
Press
Done
Refer to following figure to make each state for the
standard measurement.
Open
Load
Short
Insert short bar
(equipped with
fixture)
Nothing
Insert 50 ohm
resister
For load state, use leaded 50 ohm standard (5012-
8846) equipped with E5061B-720
4. Connect DUT then scale measurement trace
a. Connect DUT to the fixture
b. Press [Scale] then press Auto Scale All
5. Perform equivalent circuit analysis
a. Select equivalent circuit
Press [Analysis] then press Equivalent Circuit
Press Select circuit then press D
Press Display to turn on analysis
Press
Calculate
Equivalent circuit is calculated
with the specified circuit model
and measured impedance value.
b. Perform simulation
Press [Analysis] then press Equivalent Circuit
Press Simulate
Press C1 then change value by [
↑
] or [
↓
]
Measured trace and
simulated trace can be
compared by changing
each component’s value.
Simulated trace is stored
in the Memory trace.
Which is the right circuit model for my DUT?
Models for equivalent circuit should be selected
depending on the type of the DUT.
A. Inductors with high core loss
B. Inductors with high winding resistance or
resistors with low resistance value
C. Resistors with high resistance value
D. Capacitors
E. Resistors
Measured
Simulated