Keysight E5055A SSA-X Signal Source Analyzer
15
Installation
Environmental Requirements
requirements of the safety standard. Furthermore, under that temperature
requirement, the analyzer still conforms to the requirements of the safety
standard even when placing the analyzer with cooling clearance as follows:
Protection Against Electrostatic Discharge (ESD)
Set up a static-safe work-station to protect the electronic components from
the damage by the electrostatic discharge (ESD) as shown in
Figure 1-1
Example of the static-safe work station
Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of
Power Cable in Case of Emergency
As described in
“Disconnection from Supply Source” on page 19
supply is disconnected by removing the power cable’s connector plug from
either the AC outlet or the E5055A unit. When installing the E5055A, ensure
that there is sufficient free space around the unit to permit quick disconnection
of the plug (from AC outlet or E5055A unit) in case of emergency.
Requirements
Back
≥
180 mm
Sides
≥
60 mm (both right and left)