Troubleshooting
System Recovery
Procedure to execute the factory recovery
This section describes how to return the contents of the C drive to the factory
state.
Step 1.
Shut down the E5055A.
Step 2.
Disconnect all of the USB device from the USB ports.
Step 3.
Connect the keyboard to the E5055A.
Step 4.
Press the standby switch of the E5055A to turn it ON.
Step 5.
When the windows boot manager screen appears, select
Instrument Image
Recovery System
and press
[Enter]
.
Step 6. Type 2 to select Recover the original factory system image option and press
[Enter].
Figure 2-1
System recovery selection screen
You need a keyboard for this operation.
After several seconds, the next screen appears automatically even if you
do not press any key.