Chapter 4
55
Troubleshooting
Function Specific Troubleshooting
4. T
roub
lesho
o
ting
Function Specific Troubleshooting
If the E4981A exhibits a failure symptom that is related to a specific function or control
such as a front panel key control, display, data storage, remote control interface, external
trigger, isolate the trouble using the Function Specific Troubleshooting procedures
described below. The major functions of the E4981A and the troubleshooting procedure for
each function are shown in
Table 4-4
Major functions and troubleshooting procedures
Function
Description
Troubleshooting
Front panel keys
All the E4981A functions can be set and controlled
via the front panel keys.
Refer to
LCD
Almost all the information including the
measurement value, setup state, result data
processing, menu bar, softkey label and others are
indicated on the LCD.
Refer to
Front USB port
The USB port is used to save data in a USB
memory.
Refer to
.
External trigger input
The external trigger input terminal (BNC) on the
rear panel allows an external trigger source to be
used for measurement trigger.
Refer to
External Trigger Input” on
page 57
.
LAN port
The LAN interface port on the rear panel allows
the E4981A to be connected to 10/100 Base-T
Ethernet.
Refer to
.
GPIB Interface
The GPIB compatibility allows the E4981A to be
operated as a talker/listener on IEEE 488 interface
bus.
Refer to
USB (USBTMC)
Interface
Through this port, you can control the E4981A via
external controllers. It is equivalent to control via
GPIB.
USB (USBTMC) Interface”
on page 60
Handler interface
The interface is for data exchange with an
automatic machine (handler) used on production
line.
Handler Interface Function”
on page 60
.
Scanner interface
The interface to connect a scanner to perform up to
128 sets of multi-channel correction and
measurements.
Scanner Interface Function”
on page 64
.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...