:TEST:HANDler:STATus:EOM
Syntax
:TEST:HANDler:STATus:EOM {ON|OFF|1|0}
Description
This command sets the Handler End of Measurement (EOM) status for test purpose and
can be used to check/troubleshoot the Test pin signal of Handler interface. (No Query)
NOTE
is set to ON, the interface pin signal value can be
controlled/read by :TEST:HAND:xxxx commands. After completing the test,
should set to OFF so that handler interface pin signal is generated
according to the actual measurement result.
Parameters
Related commands
:TEST:HANDler:STATus:ALARm on page 286
:TEST:HANDler:STATus:INDex on page 288
:TEST:HANDler:STATus:NC on page 289
:TEST:HANDler:STATus:OVLD on page 290
:TEST:HANDler:STATus:RDYTrig on page 291
Equivalent key
sequence
No equivalent key is available on the front panel.
Description
ON
Sets Handler (End of Measurement) EOM status to ON.
OFF
Sets Handler (End of Measurement) EOM status to OFF.