:SYSTem:HANDler:TRIGger:VOLTage
Syntax
:SYSTem:HANDler:TRIGger:VOLTage <numeric>
:SYSTem:HANDler:TRIGger:VOLTage?
Description
Sets the handler trigger input voltage.
Parameters
NOTE
The value of this command is not changed with
and is only
changed with Factory Default reset (available only through Front Panel).
CAUTION
The product may get damaged if incorrect voltage is set to Handler Trigger. Take caution
while changing the Handler Trigger voltage.
Query response
{numeric}<newline><^END>
Equivalent key
No equivalent key is available on the front panel.
<numeric>
Description
The value of the handler trigger input voltage
Range
5 to 24
Initial (Factory
set) value
24
Unit
V (voltage)
Resolution